10.8V 6.4Ah 18650 Lithium ion Battery for Medical Ventilator

Additional information

Rated Voltage

10.8V

Rated Capacity

6400mAh

Cell model

INR18650-35E/3350mAh/3.6V

Battery Combination

10.8V-6.4Ah

Charging voltage

12.6V

Charging current

9V

Charging temperature

0~45℃

Discharge temperature

-20~55℃

Weight

400g

Size

154.65×59×22.2mm

Battery shell

industrial grade

Product Details

Lithium battery protection: short circuit protection, overcharge protection, over-discharge protection, over current protection, equalization, with SMBus communication, etc. We are happy to customize the charging and discharge current rate based on customer demands.

Precautions for li ion 18650 battery in series and parallel

  • Do not use batteries of different brands together.
  • Do not use batteries of different voltages together.
  • Do not mix different capacities or old and new lithium batteries.
  • Batteries with different chemical materials cannot be mixed, such as NiMH and Lithium batteries.
  • When the battery power is low, replace all batteries.
  • Use the lithium battery protection board with the corresponding parameters.
  • Choose batteries with the same performance. Generally, lithium batteries need to be paired in series and parallel. Matching standard: voltage difference≤10mV, internal resistance difference≤5mΩ, capacity difference≤20mA

18650 lithium rechargeable battery of different voltages are connected in series

Due to the consistency problem of lithium batteries, if they are connected in series in the same system (such as ternary or iron-lithium), it is also necessary to select those with the same voltage, internal resistance and capacity for grouping. Batteries with different voltage platforms and different internal resistances are used in series, which will cause a certain battery to be fully charged first and then discharged in each cycle.

If there is a protection board and there is no failure, the capacity of the whole group will be reduced. Damage to the battery caused by overcharging or overcharging.

Parallel connection of li ion 18650 of different capacities

If different capacities or old and new lithium batteries are used together, there may be liquid leakage, zero voltage and other phenomena. This is due to the difference in capacity during the charging process, which causes some batteries to be overcharged during charging, and some batteries are not fully charged. When discharging, some batteries with high capacity are not fully discharged, while those with low capacity are over-discharged. Such a vicious cycle, the battery gets damaged and leaks or has low (zero) voltage.

18650 battery pack charging precautions

  • Lithium batteries must use special chargers, otherwise they may not reach saturation and affect their performance.
  • Lithium batteries do not need to be fully discharged before charging.
  • Do not leave the charger plugged into the socket for a long time, and remove the battery from the charger as soon as possible after the battery is fully charged.
  • The battery should be removed from the electrical appliance that will not be used for a long time, and the battery should be emptied and stored.
  • Do not insert the positive and negative terminals of the battery in reverse, otherwise the battery will swell or rupture.
  • Nickel chargers and lithium chargers cannot be mixed.

Ask For A Quick Quote

We will contact you within 1 working day, please pay attention to the email with the suffix “@hlcbattery.com”